| Availability: | |
|---|---|
| Quantity: | |
MCL1808
MECANMED
Model: MCL1808
The MECANMED Fully Automated Dry Chemistry Analyzer streamlines medium-to-high throughput testing for central labs and hospitals. It delivers efficient, reliable results through full automation and a precise optical system, tackling daily batches of 50-100 samples with ease.

High-Throughput Automation: Processes ≥400 tests/hour with an open-reagent system for flexible operation.
Precision & Reliability: Features a 340-800nm optical system and a stable 37.0°C ± 0.1°C water bath for accurate results.
Robust & Easy to Use: Designed for reliable performance with minimal maintenance, featuring an intuitive interface.
Parameter | Specification |
Throughput | ≥400 tests/hour (Colorimetric) |
Sample Handling | Volume: 1.5-35 µl; Accuracy: CV ≤1.0% |
Test Menu Flexibility | ≥50 simultaneous online test items |
Reagent System | Open channels; 20ml / 70ml bottles |
Optical System | Wavelength: 340-800 nm; Absorbance: 0-3.0A |
Incubation | 37.0°C ± 0.1°C (Circulating water bath) |
Data Management | Storage: ≥500,000 results; Interface: RS-232C |
Operating Environment | Temperature: 15 - 32°C |
Central Laboratory & Hospital Core Lab: Efficiently handles high-volume, diverse testing menus.
Mobile Clinic & Satellite Labs: Its robust design supports reliable operation in various settings.
Q: What are the benefits of a MECANMED Dry Chemistry Analyzer for a Hospital?
A: The MECANMED Dry Chemistry Analyzer offers walk-away automation and high throughput (≥400 tests/hour), ideal for a Hospital's fast-paced environment.
Q: Why choose a Portable Chemistry Analyzer for a Mobile Clinic?
A: The MECANMED Portable Chemistry Analyzer is compact and designed for point-of-care use in a Mobile Clinic, providing essential diagnostics on-site.
Q: How easy is it to operate the MECANMED Chemistry Analyzer?
A: It is designed for intuitive use. The fully automated process minimizes training time, making it suitable for labs with high staff turnover.
Q: How does it ensure accuracy in high-temperature areas?
A: Its precise circulating water bath maintains a stable 37.0°C reaction temperature, ensuring consistent results despite external heat
Model: MCL1808
The MECANMED Fully Automated Dry Chemistry Analyzer streamlines medium-to-high throughput testing for central labs and hospitals. It delivers efficient, reliable results through full automation and a precise optical system, tackling daily batches of 50-100 samples with ease.

High-Throughput Automation: Processes ≥400 tests/hour with an open-reagent system for flexible operation.
Precision & Reliability: Features a 340-800nm optical system and a stable 37.0°C ± 0.1°C water bath for accurate results.
Robust & Easy to Use: Designed for reliable performance with minimal maintenance, featuring an intuitive interface.
Parameter | Specification |
Throughput | ≥400 tests/hour (Colorimetric) |
Sample Handling | Volume: 1.5-35 µl; Accuracy: CV ≤1.0% |
Test Menu Flexibility | ≥50 simultaneous online test items |
Reagent System | Open channels; 20ml / 70ml bottles |
Optical System | Wavelength: 340-800 nm; Absorbance: 0-3.0A |
Incubation | 37.0°C ± 0.1°C (Circulating water bath) |
Data Management | Storage: ≥500,000 results; Interface: RS-232C |
Operating Environment | Temperature: 15 - 32°C |
Central Laboratory & Hospital Core Lab: Efficiently handles high-volume, diverse testing menus.
Mobile Clinic & Satellite Labs: Its robust design supports reliable operation in various settings.
Q: What are the benefits of a MECANMED Dry Chemistry Analyzer for a Hospital?
A: The MECANMED Dry Chemistry Analyzer offers walk-away automation and high throughput (≥400 tests/hour), ideal for a Hospital's fast-paced environment.
Q: Why choose a Portable Chemistry Analyzer for a Mobile Clinic?
A: The MECANMED Portable Chemistry Analyzer is compact and designed for point-of-care use in a Mobile Clinic, providing essential diagnostics on-site.
Q: How easy is it to operate the MECANMED Chemistry Analyzer?
A: It is designed for intuitive use. The fully automated process minimizes training time, making it suitable for labs with high staff turnover.
Q: How does it ensure accuracy in high-temperature areas?
A: Its precise circulating water bath maintains a stable 37.0°C reaction temperature, ensuring consistent results despite external heat