PRODUCT DETAIL
You are here: Home » Products » Hemodialysis » Hemodialysis Machine » On-line HDF Hemodialysis Machine

loading

On-line HDF Hemodialysis Machine

This machine features state-of-the-art technology, including HD Hemodialysis Dialyzer, to provide superior filtration and purification. Improve patient outcomes with the latest in hemodialysis technology.
Availability:
Quantity:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
whatsapp sharing button
sharethis sharing button
  • MCX0022

  • MeCan

On-line HDF Hemodialysis Machine

Model:MCX0022

 

 

The HD Hemodialysis Machine is a state-of-the-art medical device designed for effective renal replacement therapy. Engineered with advanced technology and precision, this machine provides efficient dialysis treatment in both clinical and home settings, ensuring optimal patient outcomes and comfort during therapy.

 Haemofiltration Machine1MCX0022:Haemofiltration Machine picture (3)MCX0022:Haemofiltration Machine picture (4)MCX0022:Haemofiltration Machine picture (5)Haemofiltration Machine

Main performance of this device:

1. On-line HDF

2. Self-checking function;

3. Carbonate dialysis;

4. Double needle dialysis;

5. Liquid level detector;

6. Bubble detector;

7. Blood leakage detector;

8. Temperature and electric conductivity monitoring;

9. Arterial pressure, venous pressure and transmembrane pressure monitoring;

10. Rolling blood pump;

11. Heparin pump;

12. Dehydration amount controlled by capacity;

13. Automatic disinfection cleaning program;

14. Stand-by power of blood pump in case of power failure;

15. Information display function of display screen.

 

Parameters:

Technical Parameter

Size &Weight:

Size: 380mmx400x1380mm (L*W*H);

Net Weight approx.: 88KG;

Gross Weight approx.: around 100KG;

Package Size approx.: 650×690×1581mm (L x W x H)

Power supply:

AC220V, 50Hz/60Hz, 10A

Input power: 1500W

Back-up battery: 30 minutes

Working Condition:

Water input pressure: 0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I

Water input temperature: 5~30

Working environment temperature: 10~30at relative humidity 70%

UF rate:

Flow range: 0ml/h~4000ml/h

Resolution ratio: 1ml

Precision: ±30ml/h

Blood pump & substitution pump:

Blood pump flow range: 10ml/min~600ml/min (diameter: 8mm or 6mm)

Substitution pump flow range: 10ml/min~300ml/min (diameter 8mm or 6mm)

Resolution ratio: 0.1ml

Precision: ±10ml or 10% of reading

Heparin pump:

Syringe size: 20, 30, 50ml

Flow range: 0ml/h~10ml/h

Resolution ratio: 0.1ml

Precision: ±5%

Monitoring system &Alarm setup:

Venous pressure: -180mmHg ~ +600mmHg, ±10mmHg

Arterial pressure: -380mmHg ~ +400mmHg, ±10mmHg

TMP: -180mmHg ~ +600mmHg, ±20mmHg

Dialysate temperature: preset range 34.0~39.0

Dialysate flow: Less than 800 ml/min (Adjustable)

Substitution flow range: 0-28 L/H (on line HDF)

Blood leak detection: Photo chromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate.

Bubble detection: Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow

Conductivity: Acoustic-optic

Disinfection/Sanitize:

1. Hot disinfection:

Time: 30 minutes; Temperature: about 80, at flow rate 500ml/min;

2. Chemical disinfection:

Time: 30minutes, Temperature: about 36~50, at flow rate 500ml/min;

3. Chemical disinfection with heat:

Time: 45minutes, Temperature: about 36~80, at flow rate 50ml/min;

4. Rinse:

Time: 10minutes, Temperature: about 37, at flow rate 800ml/min;

Storage Environment:

Storage temperature should be between 5~40, at relative humidity 80%

Function:

HDF, on-line BPM, Bi-cart and 2 pcs endotoxin filters

Optional function:

Online KT/V, LAN