PRODUCT DETAIL
You are here: Home » Products » Hemodialysis » Hemodialysis Machine » Technical Data HD Hemodialysis Machine

loading

Technical Data HD Hemodialysis Machine

Engineered with advanced technology and precision, this machine provides efficient dialysis treatment in both clinical and home settings, ensuring optimal patient outcomes and comfort during therapy.
Availability:
Quantity:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
whatsapp sharing button
sharethis sharing button
  • MCX0021

  • MeCan

Technical Data HD Hemodialysis Machine

Model:MCX0021

  

The HD Hemodialysis Machine is a state-of-the-art medical device designed for effective renal replacement therapy. Engineered with advanced technology and precision, this machine provides efficient dialysis treatment in both clinical and home settings, ensuring optimal patient outcomes and comfort during therapy.

MCX0021:Hemodialysis Machine picture (3)

 

More Details:

MCX0021:Hemodialysis Machine picture (6)MCX0021:Hemodialysis Machine picture (8)Hemodialysis in Nigeria (1)

Hemodialysis in Nigeria (2)

Hemodialysis in Uganda

Main performance of this device:

1. Self-checking function;

2. Carbonate dialysis;

3. Double needle dialysis;

4. Liquid level detector;

5. Bubble detector;

6. Blood leakage detector;

7. Temperature and electric conductivity monitoring;

8. Arterial pressure, venous pressure and transmembrane pressure monitoring;

9. Rolling blood pump;

10. Heparin pump;

11. Dehydration amount controlled by capacity;

12. Automatic disinfection cleaning program;

13. Stand-by power of blood pump in case of power failure;

14. Information display function of display screen.

 

Parameters:

General data:

Dimensions :1380 x 380 x 400 mm (H x W x D);

Weight approx. :88 KG;

Water supply:

Water inlet pressure :1.5-6.0 bar

Water inlet temperature :5~32

Max. drain height :1m

Concentrate supply

Supply pressure : 0 to 2 m suction height

Electrical data

Power supply:AC220V±10%, 47-63HZ

Current consumption:max. max. 10 A

Extracorporeal circuit

Arterial pressure monitoring

Display range: -380 mmHg to +400 mmHg

Accuracy :±10 mmHg

Resolution :1 mmHg

Venous pressure monitoring

Display range: -180 mmHg to +600 mmHg

Accuracy :±10 mmHg

Resolution :1 mmHg

Transmembrane pressure monitoring

Display range :-180 mmHg to +600 mmHg

Accuracy :±20 mmHg

Resolution :1 mmHg

Arterial blood pump

Blood flow range :30 to 600ml/min in 8 mm bloodline systems

Accuracy :±10%

Air bubble pump

By ultrasound transmission, additional optical monitoring in venous clamp.

Heparin pump:

Delivery range: 0ml/h to 10ml/h

Bolus function max. :20 mL per bolus

Syringe size :10, 20, 30, 50ml

Dialysis fluid circuit

Dialysis fluid flow range

Selectable :100-800 mL/min

Dialysis fluid temperature

Selectable :34.0to 40.0

Dialysis fluid conductivity

Range: 12 to 16 mS/cm (25)

Accuracy :±0.1 mS/cm

Acid concentration dialysis fluid

Default mixing ratio :1+34 (others possible)

Range :125 to 150 mmol/L

Bicarbonate concentration dialysis fluid

Default mixing ratio :1+27.6 (others possible)

Range :-8 to +8 mmol/L bicarbonate

Ultrafiltration

UF rate :0ml/h~4000ml/h

Accuracy :±1%

Allowed dialyzer UF factor :unlimited

Parameter displayed :UF goal, UF time, UF rate, UF volume

Blood leak detector

Sensitivity :0.35 mL blood/min (HCT=32)

KT/V

Calculation

Disinfection and cleaning programs

Rinse

Temperature/flow :30 to 60/ 300 to 800ml/min

Hot Rinse (recirculation)

Temperature/flow :30 to 85/ 300 to 800ml/min

Hot disinfection with chemical

Temperature/flow :30 to 85/ 300 to 800ml/min

Various program combination selectable

Technical Data-Options

Blood pressure Monitor (BPM) (optional function)

Display range Systole: 40-280 mmHg

Diastole: 40-280 mmHg

Accuracy: 1 mmHg

Central Delivery system for acid conc. 

Supply pressure :0 to 100 mbar, 1 m suction height with Central delivery system: 0-500 mbar

Endotoxin filter-- Dialysis fluid filter system (optional function)

Balancing accuracy :±0.1% of dialysate flow

Bicarbonate dry

Concentrate :Bi-cart (optional function)